Beurer GL32 mg/dL
23
Expected results
At room temperature, the measurement results for 95% of all control solution tests should
lie within the results range printed on the test strip container.
WARNING
The results range printed on the test strip container applies only for the control
solution. This is not a recommended value for your blood sugar level.
If measurement results are outside the specified range, check the following possible
causes:
Cause Remedy
The drop of control solution was not
discarded.
The tip of the bottle was not wiped
clean.
The bottle was not shaken vigourously
enough.
Eliminate the cause and repeat the test.
Control solution has expired or is con-
taminated.
Repeat the test using a new bottle of
control solution.
The control solution, test strip, or measur-
ing device is too warm or too cold.
Allow the measuring device, test strips
and control solution to reach room
temperature (+20°C to +25°C) and
repeat the test.
Test strip is damaged. Repeat the test using a new test strip.
Test strip is out of date. Open a new container of test strips.
Repeat the test.
There is a problem with the measuring
device.
Contact customer service.
WARNING
If you persistently obtain results outside of the specified range when testing
with the control solution, do not continue using the system to determine
your blood sugar content. Contact customer service.